<5. CONFIGURATION>

5-5

IM 11M12D01-61E

Figure 5.3 Function Block and Communications Scheduling Example.

When the control period (macrocycle) is set to more than 4 seconds, set the following interval to be

more than 1% of the control period.

- Interval between “end of block execution” and “start of sending CD from LAS”,

(refer to Sec. A6.1).

- Interval between “end of block execution” and “start of next block execution”.

5.4 Setting of Tags and Addresses

This section explains the procedure for setting AV550G PD tags and node addresses. As shown in

Fig. 5.4, Fieldbus devices may be in any of three states, but only in SM_OPERATIONAL state (at the

bottom) can the function block be executed. If you are changing the PD tags and/or node address of

an AV550G, be sure to revert it to this state.

Fig. 5.4 PD-tag and Node-Address related State Transition diagram (figure reference)
Unless specified otherwise at order time, factory default settings for the AV550G are for PD tag setting

of “AV550G” and node address of F7 hex (247). To change just the node address setting, you can

clear the existing node address then set a new one; but to change the PD tag then you need to clear

the node address and PD tag the set the new PD tag and the node address.
A device whose node address has been cleared will select an arbitrary address in the range F8 to

FB (248 to 251). To address such a device, you should specify the device ID. For the AV550G, this is

5945430401xxxxxx (six alphanumeric digits after the 5945430401).

FI103

FI100

FC100

FIC100

FC200

FIC200

FC100

FI200

FI200

Function

Block

Schedule

Commu-

nication

Schedule

OUT

IN

OUT

CAS_IN

BKCAL_OUT

BKCAL_IN

BKCAL_IN

BKCAL_OUT

IN

Scheduled

Communication

F0503.ai

Macrocycle (Control Period)

Communications

need not be scheduled

UNINITIALIZED

(Neither tag nor address is set)

Tag clear

Tag setting

INITIALIZED

(Only tag is set)

SM_OPERATIONAL

(Tag and address are defined, so

the function block can be executed.)

Address clear

F0504.ai

Address setting