<5. CONFIGURATION>

5-6

IM 11M12D01-61E

5.5 Communication Setting

To set the communication function, it is necessary to change the database residing in SM (System

Management)-VFD.

5.5.1 VCR Setting

The VCR (Virtual Communication Relationship) specifies the device and item communicated with,

and the type of resources used. Each AV550G supports 33 VCRs; the first is used for management,

the other 32 are user-customizable.
The AV550G supports the following four types of VCR:
Server (QUB) VCR

A server responds to requests from, and is used for exchanging data with, a host.

This is called a QUB (Queued User-triggered Bidirectional) VCR.

Source (QUU) VCR

A source multicasts alarms or trends to other devices.

This is called a QUU (Queued Usertriggered Unidirectional) VCR.

Publisher (BNU) VCR

A publisher multicasts outputs of AI blocks, DI blocks, and MAI blocks to other function blocks.

This is called a BNU (Buffered Networktriggered Unidirectional) VCR.

Subscriber (BNU) VCR

A subscriber connects outputs from other function blocks to an MAO block.

Each VCR has the parameters listed in Table 5.4.

Parameters must be changed together for each VCR because modification for each parameter may

cause a contradiction.

Table 5.4 VCR Static Entry

T0504E-1.ai

Sub-

index

Parameter

Description

1

FasArTypeAndRole

Indicates the type and role of communication (VCR).

The following 4 types are used for the AV550G.

0x32: Server (Responds to requests from host.)

0x44: Source (Transmits alarm or trend.)

0x66: Publisher (Sends AI, DI, MAI block output to other blocks.)

0x76: Subscriber (Connects output of other blocks to MAO block.)

2

FasDllLocalAddr

3

FasDllConfigured

RemoteAddr

Sets the node address of the called party for communication and

the address (DLSAP or DLCEP) used to specify VCR at that

address. For DLSAP or DLCEP, a range of 20 to F7 in

hexadecimal is used. Addresses in Subindex 2 and 3 need to be

set to the same contents of the VCR as the called party (local and

remote are reversed).

Sets the local DLSAP or DLCEP address to specify a VCR in the

AV550G. A range of 20 to F7 in hexadecimal.

4

FasDllSDAP

Specifies the quality of communication. Usually, one of the

following types is set.

0x2B: Server

0x01: Source (Alert)

0x03: Source (Trend)

0x91: Publisher/Subscriber

5

FasDllMaxConfirm

DelayOnConnect

To establish connection for communication, a maximum wait time

for the called party's response is set in ms. Typical value is 60

seconds (60000).