<6. IN-PROCESS OPERATION>

6-1

IM 11M12D01-61E

6. IN-PROCESS OPERATION

This section describes AV550G function block mode transitions and status changes during operation.

6.1 Mode Transition

If the function block mode is changed (from Auto) to O/S (Out_Of_Service), the function block is

stopped and a block alarm is generated.
If the function block mode is changed (from Auto) to Manual, the function block stops updating output

values. In this case it is possible to write desired output values to the OUT parameter.

However, the status of parameters cannot be changed.

6.2 Generation of Alarm or Error Status

6.2.1 AV550G Alarms

If the AV550G self-diagnostic functions detect an abnormality in the AV550G, a device alarm is

generated by the resource block. If an abnormality (block error) or process (value abnormal) alarm is

detected, the corresponding block generates an alarm.

6.2.2 Alarms and Events

An AV550G can report the following alarms or events as alerts:
Analog Alerts (Generated when a process value exceeds threshold)

By AI Block:

Hi-Hi Alarm, Hi Alarm, Low Alarm, Low-Low Alarm

Discrete Alerts (Generated when an abnormal condition is detected)

By Resource Block:

Block Alarm, Write Alarm

By Transducer Block: Block Alarm

By AI Block:

Block Alarm

By DI Block:

Block Alarm

By MAI Block:

Block Alarm

By MAO Block:

Block Alarm

Update Alerts (Generated when a important (restorable) parameter is updated)

By Resource Block:

Update Event

By Transducer Block: Update Event

By AI Block:

Update Event

By DI Block:

Update Event

By MAI Block:

Update Event

By MAO Block:

Update Event

An alert has the following structure: