<8. GENERAL SPECIFICATIONS>
8-1
IM 11M12D01-61E
8. GENERAL SPECIFICATIONS
8.1 Standard Specifications
For items other than those described below, refer to GS 11M12D01-01E.
Applicable Models
All the models of AV550G with Fieldbus communication functions (Output code: F).
These models conform to the following EMC standards:
EN61326
AS/NZS2064
Output Signals
Digital communication signal compliant with the FOUNDATION Fieldbus protocol
Physical Layer Type
113 (standard power signaling, bus powered, non I.S.)
Supply Voltage
9 to 32 V DC for general-purpose, flameproof types and Nonincendive.
Condition of Communication Line
Supply voltage: 9 to 32 V DC
Supply current: 15 mA (maximum)
Functional Specifications
The communication specifications conform to the H1
fieldbus specification of the Fieldbus FOUNDATION.
Function blocks
• Three AI function blocks.
• Two DI function blocks.
• One MAI block.
• One MAO block.
Link master functionality (BASIC of factory setting) Software download function
8.2 Model and Suffix Codes
1. Detector
Refer to GS 11M12A01-01E for a detailed explanation of the detector specifications and available
accessories.