9. PARAMETER LISTS
T0903-5.EPS
Index
Parameter
Valid Range
Initial Value
Description
Write
Mode
111
DISPLAY_SELECT1
This parameter is used in order to set the LCD display mode.
Auto
1: Out Value
0 to 4
112
DISPLAY_SELECT2
Auto
0: Off
0 to 9
113
DISPLAY_SELECT3
Auto
0: Off
0 to 9
114
DISPLAY_CYCLE
Auto
1: 400ms
This parameter is used in order to set the renewal cycle of LCD
display.
Renewal cycle: 200 msec “DISPLAY_CYCLE”
0: 200ms
1: 400ms
2: 1s,
3: 2s,
4: 4s,
5: 8s
115
LANGUAGE
Auto
0: English
0: English
116
RATE_LIMIT
O/S
5
This parameter is used in order to set the rate limit value.
0 to 10
117
DEAD_TIME
O/S
0
This parameter is used in order to set the time for application of the
rate limit, and if a value of 0 is set, the rate limit function will be
terminated.
0 to 15
118
PULSING_FLOW
O/S
0: No
This parameter is used in order to set the pulsing flow support.
0: No
1: Yes
119
POWER_SYNCH
O/S
1: Yes
This parameter is used in order to indicate whether or not the
internal frequency is to be synchronized with that of the power supply.
0: No
1: Yes
120
POWER_FREQUENCY
O/S
50
This parameter is used in order to set the power frequency.
47.00 to 63.00
121
SOFTWARE_REV_NO
–
This parameter is used to display the software revision number.
122
ADHESION_CHECK
O/S
0: No
This parameter selects whether or not the adhesion diagnostic
function will be carried out.
NOTE
If the judgment value for Level 3 is exceeded, a warning is displayed;
and if the value for Level 4 is exceeded, an alarm is displayed.
0: No
1: Yes
123
ADHESION_LEVEL1
O/S
0.10
This parameter is used in order to set the resistance value for
judgment of Level 1.
0.00 to 100.00
124
ADHESION_LEVEL2
O/S
0.50
This parameter is used in order to set the resistance value for
judgment of Level 2.
0.00 to 100.00
125
ADHESION_LEVEL3
O/S
1.00
This parameter is used in order to set the resistance value for
judgment of Level 3.
0.00 to 100.00
126
ADHESION_LEVEL4
O/S
3.00
This parameter is used in order to set the resistance value for
judgment of Level 4.
0.00 to 100.00
127
ADH_MEASURE_VALUE
–
–
This parameter displays the value measured using the adhesion
diagnostic function.
0.00 to 1000.00