Yokogawa AXFA14G/C User Manual
Page 21

IM 01E20F12-01E
4-7
4. CONFIGURATION
Table 4.4
TOTAL
Quality (bit 6-7)
Sub-status (bit 2-5)
Limit (bit 0-1)
Alarm Value
T0404.EPS
2:Good (NC)
0:Ok
0:ok
1:Update Event
2:Active Advisory Alarm
1:low limited
--:Total1 Lo Alm
0x89
--:Total2 Lo Alm
--:Total3 Lo Alm
2:high limited
--:Total1 Hi Alm
0x8A
--:Total2 Hi Alm
--:Total3 Hi Alm
3:Active Critical Alarm:
1:low limited
--:Total1 Lo Lo Alm
0x8D
--:Total2 Lo Lo Alm
--:Total3 Lo Lo Alm
2:high limited
--:Total1 Hi Hi Alm
0x8E
--:Total2 Hi Hi Alm
--:Total3 Hi Hi Alm
4:Unack. Update Event
5:Unack. Advisory Alarm
6:Unack. Critical Alarm
9:Maintenance required
0:ok
80:Adhesion Wng
0xA4
81:(Reserved)
1:Uncertain
0:Non specific
0:ok
10:uP Fault
0x40
11:EEPROM Fault
12:A/D(H) Fault
13:A/D(L) Fault
14:A/D(Z) Fault
16:EEPROM Dflt
100:Comm uP Fault
101:Comm EEPROM Fault
105:Comm Error1
106:Comm Error2
15:Coil Open
30:Sig Overflow
31:Empty Pipe
33:Adhesion Alm
34:(Reserved)
41:TB O/S Mode
50:Span > 10m/s
51:Span < 0.1m/s
57:Dens Set Err
71:Meas Mod Set
72:Size Set Err
73:Adh Set Err
82:Auto Zero Wng
1:low limited or
85:Flow Vel Over
0x41
2:high limited
0x42
0:Bad
1:Configuration Error
0:ok
120:Total1 Unit Set Err
0x04
121:Total2 Unit Set Err
122:Total3 Unit Set Err
7:Out of Service
3:constant
40:PB O/S Mode
0x1F
43:Total1 FB O/S Mode
44:Total2 FB O/S Mode
45:Total3 FB O/S Mode