4. CONFIGURATION
DISPLAY_SELECT1,2,3:
Table 4.2
DISPLAY_SELECT1, 2, 3
DISPLAY_SELECT 1
DISPLAY_SELECT 2
DISPLAY_SELECT 3
The display content for the display unitís
first line.
The display content for the display unitís
second line.
The display content for the display unitís
third line.
0:Flow Rate(%)
1:Out Value
2:Totalized Flow Rate1
3:Totalized Flow Rate2
4:Totalized Flow Rate3
The factory default setting is 1.
0:Off
1:Flow Rate(%)
2:Out Value
3:Flow Rate(Bar)
4:Totalized Flow Rate1
5:Totalized Flow Rate2
6:Totalized Flow Rate3
7:Tag No
8:Adhesion Check
9:Communication
The factory default setting is 0.
0:Off
1:Flow Rate(%)
2:Out Value
3:Flow Rate(Bar)
4:Totalized Flow Rate1
5:Totalized Flow Rate2
6:Totalized Flow Rate3
7:Tag No
8:Adhesion Check
9:Communication
The factory default setting is 0.
T0402.EPS
DISPLAY_CYCLE:
Specifies the cycle of LCD display. The factory default
setting is 400ms, but if a low temperature environment
makes it difficult to view the display, it is recom-
mended that you set a longer display cycle.
VOLUME_FLOW:
Indicates the current measured value and status as the
primary value (volumetric flow) This parameter is
input to the AI Function Block and the Totalizer
Function Block.
Table 4.3
VOLUME_FLOW
Quality (bit 6-7)
Sub-status (bit 2-5)
Limit (bit 0-1)
Alarm Value
T0403.EPS
2:Good (NC)
0:Ok
0:ok
Normal condition
1:Update Event
9:Maintenance required
0:ok
80:Adhesion Wng
0xA4
81:(Reserved)
1:Uncertain
4:Sensor Conversion Not Accurate
1:low limited or
85:Flow Vel Over
0x49
2:high limited
0x4A
7:Configuration Error
0:ok
82:Auto Zero Wng
0x5C
0:Bad
0:Non specific
0:ok
40:PB O/S Mode
0x00
1:Configuration Error
3:constant
50:Span > 10m/s
0x07
51:Span < 0.1m/s
57:Dens Set Err
71:Meas Mod Set
72:Size Set Err
73:Adh Set Err
3:Device Failure
3:constant
10:uP Fault
0x0F
11:EEPROM Fault
12:A/D(H) Fault
13:A/D(L) Fault
14:A/D(Z) Fault
16:EEPROM Dflt
0:ok
100:Comm uP Fault
0x0C
101:Comm EEPROM Fault
105:Comm Error1
106:Comm Error2
4:Sensor Failure
3:constant
15:Coil Open
0x13
30:Sig Overflow
31:Empty Pipe
33:Adhesion Alm
34:(Reserved)
7:Out of Service
0:ok
41:TB O/S Mode
0x1C