Table 6.3.3 Alarm parameters as simulation function in the Physical block
T0605.EPS
Index
Parameters
Explanation
52
DIAGNOSIS_SIM
_MODE
When the hardware simulation jumper
is on, DIAGNOSIS_SIM_MODE is
set to enable, and the bits are set to
“on”, the alarms (or warnings) can be
modified.
Definition of supported
DIAGNOSIS_EXTENSION
information-bits for testing alarms and
warnings can be modified.
53
DIAGNOSIS_SIM
_EXTENSION
Valid Range
0: Disabled,
1: Enabled
0: Off,
1: On
(Note 1)
Initial Value
0: Disabled
0x0000 0000 0000
0000 0000 0000
Note 1: When the valid range is set to “1: ON”, the following data is necessary to set.
Alarm
No.
Octet
1
Bit
0
10
Name
uP Fault
EEPROM Fault
A/D(H) Fault
A/D(L) Fault
A/D(Z) Fault
Coil Open
EEPROM Dflt
Comm uP Fault
Comm EEPROM Fault
Total1 Not Saved
Total2 Not Saved
Total3 Not Saved
Comm Error1
Comm Error2
(Reserved)
(Reserved)
(Not Used)
(Not Used)
(Not Used)
(Not Used)
Sig Overflow
Empty Pipe
Adhesion Alm
(Reserved)
1
11
2
12
3
13
4
14
5
15
6
16
7 100
2
0 101
1 102
2 103
3 104
4 105
5 106
6 107
7 108
3
0
1
2
3
4
30
5
31
6
33
7
34
Alarm
No.
Octet
4
Bit
0
50
Name
Span > 10m/s
Span < 0.1m/s
Dens Set Err
Meas Mod Set
Size Set Err
Adh Set Err
(Not Used)
(Not Used)
(Not Used)
(Not Used)
(Not Used)
(Not Used)
Adhesion Wng
(Reserved)
Auto Zero Wng
Flow Vel Over
(Reserved)
(Not Used)
(Not Used)
(Not Used)
(Not Used)
(Not Used)
(Not Used)
Sim. Jmpr On