Yokogawa AXFA14G/C User Manual
Page 43

IM 01E20F12-01E
5-19
5. EXPLANATION OF BASIC ITEMS
Warning Message
On the Integral LCD indicator, following messages are
displayed when warning is generated.
Table 5.6.6 Warning Message Display
Category
Display Message
Warning
T0514.eps
80:Adhesion Wng
Slight adhesion to electrodes.
82:Auto Zero Wng
Results of automatic zero adjustment are higher than the rated values.
84:Disp Over Wng
Overflow in the display digits during instantaneous flow rate display.
Display at the LCD indicator only.
85:Flow Vel Over
Flow velocity overflow.
140:Sim.Jmpr On
Simulation jumper is ON.
141:PB Sim. Enabled
PB.DIAGNOSIS_SIM_MODE is enabled.
142:TB VF Sim. Enabled TB.SIMULATE_MODE is ON (Volume Flow Value).
143:AI Sim. Enabled
AI.SIMULATE is enabled.
144:Total1 Sim. Enabled
TOTAL1.SIMULATE is enabled.
145:Total2 Sim. Enabled
TOTAL2.SIMULATE is enabled.
146:Total3 Sim. Enabled
TOTAL3.SIMULATE is enabled.
147:DI1 Sim. Enabled
DI1.SIMULATE is enabled.
148:DI2 Sim. Enabled
DI2.SIMULATE is enabled.
150:AI FB Man Mode
AI.TARGET_MODE is Manual mode.
151:Total1 FB Man Mode TOTAL1.TARGET_MODE is Manual mode.
152:Total2 FB Man Mode TOTAL2.TARGET_MODE is Manual mode.
153:Total3 FB Man Mode TOTAL3.TARGET_MODE is Manual mode.
154:DI1 FB Man Mode
DI1.TARGET_MODE is Manual mode.
155:DI2 FB Man Mode
DI2.TARGET_MODE is Manual mode.
Description
Auto Zero Display
When ZERO_POINT_ADJUST is executed, the
following is displayed until auto zeroing finishes.
A
o
N
w
o
Z e r
u t o
i n
c u t
e
E x
g . . .
F0521.eps