Canbus trigger, decode, and measure – Teledyne LeCroy CANbus TD and CANbus TDM - Operators Manual User Manual
Page 46
![background image](/manuals/353658/46/background.png)
CANbus Trigger, Decode, and Measure
46
CANbus-TD-TDM-OM-E RevB
Trigger Capability
CANbus TD and TDM
# Data Bytes = 0 to 8. Data can be defined by nibble. Triggers on that
data pattern regardless of position.
Binary:
Any combination of 0,1, or X for 1-64 bits. Triggers on that data pattern
regardless of position.
Bit Rates
10, 25, 33.333, 50, 83.333, 100, 125, 250, 500, 1000 kb/s or user-defined
between 10-1000 kb/s.
Trigger Input
Any analog Channel or the EXT input.
Trigger Design
Internal to oscilloscope, settable like any other oscilloscope trigger.
Search Capability
CANbus TD and TDM
Search Options
Search for Any Frame, Any Error or Frame ID in Hexadecimal format.
Measure and Graph
Capability
CANbus TD
CANbus TDM
CAN Timing
Measurements
Not Applicable
CAN-CAN, CAN-Analog, Time@CAN,
CAN Message Bitrate.
CAN Data Extraction
Not Applicable
CAN-Value.
CAN Bus Load
Measurements
Not Applicable
CAN Bus Load %.
Graphing Functions
Not Applicable
Track, Trend and Histogram of CAN
measurements.