Reference
CANbus TD and TDM Specifications
Note: Specifications are subject to change without notice.
Definition
CANbus TD and TDM
Protocol Setup
Select Bitrate (10, 25, 33.333, 50, 83.333, 100, 125, 250, 500, 1000 kb/s
or user-defined between 10-1000 kb/s).
Decode Capability
CANbus TD and TDM
Format
Hexadecimal
Decode Setup
Threshold definition required. Default is to Percent amplitude. Select
BitRate.
Decode Input
Any analog Channel, Memory or Math trace.
Number of Decoded
Waveforms
Up to 4 buses may be decoded at one time.
In addition, zooms can be displayed (with decoded information).
Location
Overlayed over DATA waveform, on Grid.
(Use multi-grid if there is more than one decoder ON)
Visual Aid
Color Coding for FRAME, ID, DLC, DATA, CRC, Ack, Stuff Bits and Errors.
Decode information is intelligently annotated based on timebase setting.
Trigger Capability
CANbus TD and TDM
Format
Hexadecimal or Binary
Trigger Setup
Trigger on All Frames, Frame ID, ID with Data, Remote Frames, or Error
Frames.
Address (ID)
Condition Setup
Specify one Frame ID or a range of Frame IDs. Frame ID trigger can be
combined with Data.
Conditional Trigger
Setup
Conditional Frame ID and Conditional Data triggering available. Choose
from <=, <, =, >, >=, <>, in range, out of range or don’t care conditions.
Data Setup
Hexadecimal: