66
IBTracer 4X Protocol Analyzer User’s Manual
CATC
SW Version 2.3
The event automatically connects to the Counter button and then
connects to the Trigger button.
Step 5
Click the blue dot in top right corner of your selected counter.
A pop-up menu opens. This menu lets you
configure the counter.
Step 6
Choose Change Counter Value.
Step 7
Set the counter to a value of your choice,
and click OK.
6.10 Triggering External Output
The Actions window has options enabling the analyzer to send an output
signal through a BNC cable or an RS-232 cable and breakout board to an
external device such as an oscilloscope.
Cabling
To set up the analyzer for signal output, you will first need to connect the
analyzer to the external device using a BNC or RS-232 cable.
Using a BNC Cable
If using a BNC cable, connect the external device to one end of the cable
and connect the other end of the cable to the output connector marked "Ext.
2" on the back of the analyzer.