X96-2009PL is the Scintillation Detector Interface Module. This optional

11

module provides:

• 1 isolated scintillation input (pulse counter, max signal 0-12

12

V, threshold 0.6 V)

• 1 head temperature input (1 uA per deg K)

• a 4-20 mA input (Ch. 2)
• isolated power for the scintillation detector 24 V 40 mA supply

13

.

X96-2009PL3

X96-2009PL is the Scintillation Detector Interface Module. This optional

14

module provides:

• 1 isolated scintillation input (pulse counter, max signal 0-12

15

V, threshold 0.6 V)

• 1 head temperature input (1 uA per deg K)

• a 0 –10 volt input (Ch. 2)
• isolated power for the scintillation detector 24 V 40 mA supply

16

.

11

At least one detector interface module is required.

12

8.6 V nominal.

13

The power supply has the ability to control the power to the scintillation detector:

• when commanded by the CPU module,

• when the processor on the module detects a condition that could harm the scintillation detector,

• when the watchdog timer generates a reset.

14

At least one detector interface module is required.

15

8.6 V nominal.

16

The power supply has the ability to control the power to the scintillation detector:

• when commanded by the CPU module,

• when the processor on the module detects a condition that could harm the scintillation detector,

• when the watchdog timer generates a reset.