Calibrate density menu, Ref constants – Ronan X96S MASS FLOW GAUGE User Manual
Page 35
31
Ref Constants
This menu is used to view and control the reference constants used in the reference and calibration procedures. It
contains the following items:
ITEM
FUNCTION
Ref Mode
Shows and allows the user to set the pre-selected promptings and values during a
reference.
Calibr.Mode
Shows and allows the user to set the pre-selected promptings and values during a
calibration.
Ref Time
Shows and allows the user to set the number of seconds of data to collect for a
reference or calibrate sample (0 – 999 seconds)
MinRefCnt
Shows and allows the user to set the minimum raw count value to use for a reference
or calibrate sample
MaxRefCnt
Shows and allows the user to set the maximum raw count value to use for a reference
or calibrate sample
Ref Mode is one of the following:
ITEM
MEANING
Empty
The pipe will be Empty during a reference (just air; no process in the measuring area)
Water
The pipe will be filled with water during the process
Process
Actual process material to be used for reference. (Density must be provided)
Absorber
Absorber plate will be placed in the radiation path and the indicated value noted.
Calibrate Density Menu
This menu is used to access the various mass flow calibration procedures. It contains the following items:
ITEM
FUNCTION
State
Shows the state of the mass flow configuration process
Low Reference
Selecting this item takes the user to the Low Reference menu
High Calibrate
Selecting this item takes the user to the High Calibrate menu
Manual Entry
Shows and allows the user to set the calculated Loading Factor value (Same value as
in Calibrate Menu, L.F.)
Clear Ref/Cal
This item invokes method that clears the density reference and L.F. (Loading Factor)