Calibrate density menu, Ref constants – Ronan X96S MASS FLOW GAUGE User Manual

Page 35

background image

31

Ref Constants

This menu is used to view and control the reference constants used in the reference and calibration procedures. It
contains the following items:

ITEM

FUNCTION

Ref Mode

Shows and allows the user to set the pre-selected promptings and values during a
reference.

Calibr.Mode

Shows and allows the user to set the pre-selected promptings and values during a
calibration.

Ref Time

Shows and allows the user to set the number of seconds of data to collect for a
reference or calibrate sample (0 – 999 seconds)

MinRefCnt

Shows and allows the user to set the minimum raw count value to use for a reference
or calibrate sample

MaxRefCnt

Shows and allows the user to set the maximum raw count value to use for a reference
or calibrate sample

Ref Mode is one of the following:

ITEM

MEANING

Empty

The pipe will be Empty during a reference (just air; no process in the measuring area)

Water

The pipe will be filled with water during the process

Process

Actual process material to be used for reference. (Density must be provided)

Absorber

Absorber plate will be placed in the radiation path and the indicated value noted.

Calibrate Density Menu

This menu is used to access the various mass flow calibration procedures. It contains the following items:

ITEM

FUNCTION

State

Shows the state of the mass flow configuration process

Low Reference

Selecting this item takes the user to the Low Reference menu

High Calibrate

Selecting this item takes the user to the High Calibrate menu

Manual Entry

Shows and allows the user to set the calculated Loading Factor value (Same value as
in Calibrate Menu, L.F.)

Clear Ref/Cal

This item invokes method that clears the density reference and L.F. (Loading Factor)