Detector fault menu, Temp comp menu – Ronan X96S MASS FLOW GAUGE User Manual
Page 22

18
The Empty Clamp menu is used to configure the parameters associated with the mechanism that shuts off power to the detector if
the detector received more radiation than it is capable of measuring. It contains the following items:
ITEM
FUNCTION
Empty Clamp
Shows the current state of the Empty Clamp mechanism (enabled or disabled) and
allows the user to change the state.
Off Time
Shows and allows the user to change the number of seconds that power to the detector
be shut off when the empty clamp mechanism activates before turning the detector on
to see if the radiation level has dropped to a value that the detector is capable of
measuring. Typical value ranges from 15 to 30 minutes.
Rcvry Time
Shows and allows the user to change the number of seconds that power will be applied
to the detector when the empty clamp mechanism has activated to see if the radiation
level has dropped to a value that the detector is capable of measuring. Typical value
here is 30 seconds.
Min Counts
Shows and allows the user to change the number-minimum number of counts that are
used to determine that the detector is saturated (exposed to more radiation than it is
capable of measuring).
Max Counts
Shows and allows the user to change the number maximum number of counts that are
used to determine that the detector is exposed to more radiation than it is capable of
measuring.
Detector Fault Menu
The Detector Fault menu is used to configure a window in which the detector counts must fall within, in order to complete a
reference/calibration.
Temp Comp Menu
Temperature compensation in the X96S is calculated as a quadratic function
3
. If temperature compensation is to be used, the
process temperature can be captured automatically for you during Referencing by enabling Temp auto Cap. If you choose to
enter the temperature manually, the entered value should be the temperature of the process material during the Referencing
function.
Process temperature is typically read via platinum or nickel, 2 or 3-wire RTD attached to the X96S. In these cases, the X96S
needs no temperature calibration. You can optionally feed a 0-10 Volt or 4-20mA signal into the X96S from another source that
represents the range of temperature for your process. In this case, you will need to perform a Temp Ref (in the Calibration Menu)
to set the range of measurement. You will also need to define the type of temperature device you are using in the Hardware Menu,
and Proc Temp menus.
The Temp Comp menu is used to control the temperature compensation mechanism. It contains the following items:
ITEM
FUNCTION
Temp Comp
Shows the current state of the Temp Comp mechanism (enable/disable) and allows the
user to change the state.
Temp Units
Shows and allows the user to set the temperature units used by the temperature
compensation mechanism.
A Coeff
Shows and allows the user to set the “A” coefficient that relates temperature to
changes in mass flow.
B Coeff
Shows and allows the user to set the “B” coefficient that relates to temperature to
changes in mass flow.
Enter Temp
Shows and allows the user to set the process temperature at the time that the gauge was
Referenced (if Auto Cap was ‘enabled’ during Referencing).
Temp Auto Cap
This invokes a method that captures the process temperature automatically during
gauge Referencing.
3
compensated density = uncompensated density + (A Coeff* temperature delta) + (B Coeff* temperature delta
2
)