Cal remote solids, Temp reference, Loop config menu – Ronan X96S MASS FLOW GAUGE User Manual
Page 37

33
Temp Reference
This menu is used ONLY if the Hardware Menu, Proc Temp value is set for 0-10 volts or 4-2- ma
This menu is used to perform the process temperature reference procedure. It contains the following items:
ITEM
FUNCTION
Temp LoRef
This invokes a method that performs the low temp reference
Temp High Cal
This invokes a method that performs the high temp reference
Temp. Cal Low
Temperature value during the Temp Lo Ref procedure
Temp. Cal High
Temp value during the Temp High Cal procedure
Temp Cap Low
Shows the raw captured temperature counts during the low temp reference
Temp Cap High
Shows the raw captured temperature counts during the high temp reference
RTD Slope
Factory value for RTD Slope
RTD Offset
Factory value for RTD Offset
Cal Remote Solids
This menu is used to calibrate the remote input. It contains for following items:
ITEM
FUNCTION
Lo Solids Cap
Shows the user the low solids captured counts
Hi Solids Cap
Shows the user the high solids captured counts
Min Solids
Shows and allows the user to input the minimum solids value
Max Solids
Shows and allows the user to input the maximum solids value
Low Reference
This invokes a method that performs the low reference procedure
Hi Calibration
This invokes a method that performs the high calibration procedure
Loop Config Menu
This menu is used to access the primary 4-20ma loop calibration procedures. It contains the following items:
ITEM
FUNCTION
Pv is
Shows and allows the user to set the variable assigned to the primary 4-20mA current
loop
Loop test
This item invokes a method that performs a test on the primary 4-20ma current loop
Damping
Shows and allows the user to set the damping constant for the primary 4-20ma current
loop
D/A trim
This item invokes method that performs the D/A trimming of the primary 4-20ma
current loop
PV is one of the following:
ITEM
MEANING
Mass Flow
Mass Flow (example: pounds per hour)
Density
Density (example: pounds per linear foot)
Flow Rate
Flow Rate (example: feet per second)
Head Temp
Head Temperature (if Available)
Not Assigned
Blank line