51

Quick Start Reference – Calibrating Density

Step 1

Start at the Status Display

From the Status Display Screen, Press the F3 (Lock) Key to display
the Password Screen (or the Main Menus if the password is
disabled).

C

F1

F2

F3

F4

Ronan Engineering

X96S Mass Flow Gauge

Enter Password:

*

*****

All0

RST0

EXIT

ENTER

C

F1

F2

F3

F4

Ronan Engineering

X96S Mass Flow Gauge
Status Display

Mass Flow 3.05 t/hr

Density 2.5 SPG

Proc Temp 50.00 deg
Raw Counts 12000

Filt.Counts 12000

Time 04/20/05 08:05

Diagnostics

Step 2 Password Menu

With the left most character * highlighted (All digits are set
at 000000 at this point), enter in the password ‘101010’.
Press the

key 1 time to make the left most character

equal to 1. Press the

key 2 times to move to the third

character.
Press the

key 1 time to make the third character equal to

1. Press the

key 2 times to move to the fifth character.

Press the

key 1 time to make the fifth character equal to

1. Press the

F4 key (enter) to accept the password.

This will take you to the Main Menu.

Note: If the wrong password was entered, press the F1
(All0) to set all the characters back to a value of 000000.
You can begin to re-enter the password from the beginning.
Pressing the F2 (RST0) will set the individual character that
is highlighted back to the value 0.

Note: For security reasons, each character will always be
displayed as an asterisk *.