Alarms, Auto reference – Ronan X96S MASS FLOW GAUGE User Manual

Page 28

background image

24

Alarms

The Alarms menu is used to configure the parameters associated with the analog alarms. This applies to alarms 1-8 as well as the
4-20mA alarm

ITEM

FUNCTION

Source

Show and allows the user to set the source of the alarm.

Alarm Type

Shows and allows the user to set the alarm type.

Setpoint

Shows and allows the user to set the alarm set point.

Setpoint2

Shows and allows the user to set the second alarm set point.

2

Hysterisis

Shows and allows the user to set the alarm hysterisis percent.

Source is one of the following

ITEM

MEANING

Density

Uses the Density for the source of the alarm

Mass Flow

Uses the Mass Flow for the source of the alarm

Flow Rate

Uses the Flow Rate for the source of the alarm

Proc Temp

Uses the Process Temp of the detector for the source of the alarm

Head Temp

Uses the Head Temperature of the detector for the source of the alarm

Filtered Counts

Uses the Filtered Counts from the detector for the source of the alarm

4-20 mA

Uses the 4-20 mA from the detector for the source of the alarm

Alarm Type is one of the following:

ITEM

MEANING

None

Alarm not yet set

Low

Alarm when the source is equal to or lower than Setpoint

High

Alarm when the source is equal to or higher than Setpoint

Range

Alarm when the source is equal to or lower than Setpoint OR the source is equal to or
higher than Setpoint2

Auto Reference

The Auto Reference menu is used to configure the parameters associated with the Auto Reference. It contains the following
items:

ITEM

FUNCTION

Auto Ref

Shows and allows the user to Enable or Disable the Auto Reference

Ref Delay

Shows and allows the user to change the time for the Reference Delay

Auto Ref Density

Shows and allows the user to change the Reference Mass Flow to be used for the Auto
Ref Mass Flow

2

The second alarm set point is only used when the alarm type is range.