Recalibration/Proving without a Lap Pad

NOTE

: If the new value entered is out of range for the

LCR, "Error" will appear in the display . If this occurs,

press SELECT to return to the Prover Quantity display.

Repeat the process by beginning a new prover run.

When this process is complete, turn the Selector Switch

clockwise out of the Calibration Position and replace

the Switch plate.

NOTE: S tep-by-step instructions on conducting a

recalibration/proving run appear in Appendix E.

CALIBRATION MENU

101.323

CALIBRATION - SCREEN 1

Prod 1

CALIBRATION - SCREEN 2

F 70.0

CALIBRATION - SCREEN 3

27.2 ºF

CALIBRATION - SCREEN 4

101.663

CALIBRATION - SCREEN 5

Error

ERROR SCREEN

SELECT

INCREASE

“Current
Delivery

Active”

SELECT

SELECT

SELECT

SELECT

If an unacceptable

value is entered.

SINGLE POINT RECALIBRATION/PROVING
CONTINUED

Repeat the process above for each digit position until

the value for each has been selected. When the least

significant digit is entered, the k-Factor is recalculated

and the display returns to Calibration - Screen 1.

101.663