beautypg.com

Lap pad calibration & set-up, Multi-point calibration – Liquid Controls LCR Setup & Operation User Manual

Page 24

background image

24

Lap Pad Calibration & Set-up

Multi-point calibration allows a flowmeter to be used over a wide range of flow rates without experiencing a loss in

measurement accuracy dueA to the shape of the meter curve (usually at low flow rates).

After determining the inaccuracy

of the meter at a variety of flow rates and inputting that information into the LCR multi-point calibration table, accuracy

corrections are then continuously made by the LCR based on the measured flow rate.

There are two ways to obtain the

information necessary to load the multi-point calibration table:

1. Manufacturer's data supplied with the meter can be entered directly.

2. The meter can be field calibrated by proving at various flow rates.

1. MANUFACTURER'S DATA TECHNIQUE

Step 1:

Go to Screen 7 of PRODUCT CALIBRATION. From the manufacturer's data sheet showing the accuracy

at various flow rates, enter into PT^1 the flow RATE and the delivery %ERROR corresponding to the

maximum flow rate of the meter.

Step 2:

Select up to nine additional points from the manufacturer's data sheet to characterize the remainder of

the meter's accuracy curve and enter this data into PT^2, PT^3, etc., respectively, in descending order

by rate. (While it is not necessary to enter the selected points in a specific order, it is desirable to do so

because the LCR will re-sort the data and arrange it in descending order by rate when it is APPLIED.)

To enter negative %ERROR values, first enter the number, then press the "*" key to display the "-" sign.

Errors that can occur during multi-point table data entry include:

DUPLICATE FLOW RATE ERROR

The same flow rate value was entered twice.

RANGE ERROR

A %ERROR outside the ±3.0% limit was entered.

POINT SELECTION TIPS:

Because the LCR uses a piece-wise linear interpolation method of correction, it is desirable to use

more data points along segments of the accuracy curve that have the greatest curvature (typically

in the low flow rate region).

Where the meter curve already approximates a straight line (in the high flow rate region), use

fewer of the available points.

Step 3:

Set the LINEAR^ field to APPLIED and enter it. This puts the linearizing algorithm into ef fect. If the

error message ADJACENT POINTS OUT OF 0.25% RANGE is displayed, APPLIED will not t ake

effect and the points will need to be reselected to ensure that no two points adjacent to each other are

more than 0.25% apart in %ERROR.

Step 4:

Field prove the meter at the normal high flow rate of the system. Adjust the PULSES/UNIT in Screen 3

of the PRODUCT CALIBRATION menu by entering the actual PROVER QTY (see Page 19) to obtain

accuracy within tolerance limits.

MULTI-POINT CALIBRATION

PT^
1

0.00

GALLONS

PER MINUTE

0.000

Product Calibration - Screen 7

RATE

UNITS^

RATE BASE^

%ERROR