beautypg.com

Recalibration/proving without a lap pad, Single point recalibration/proving – Liquid Controls LCR Setup & Operation User Manual

Page 31

background image

31

Recalibration/Proving without a Lap Pad

The addition of two control buttons on the LCR allows for single point recalibration or proving without the need of a Lap

Pad or other data entry device. This calibration is limited to recalculating the k-Factor or single point recalibration for up

to four dif ferent product types. These two buttons can be used to select dif ferent display values as well as enter

product selections and preset values. TheSELECT button is the front and the INCREASE button is the rear as shown

in the figure below.

If a menu item can be edited, the INCREASE button will be active and can be used to increase a value or scroll through

a list of options. The SELECT button will be used to scroll through consecutive digit s across the display (from left to

right).

STEP 1: REMOVING THE LCR SWITCHPLATE FOR INITIAL SET-UP

Remove the lead seal and wire assembly from the LCR Switchplate (if previously sealed by Weights & Measures

authorities).

Remove the four screws securing the nameplate over the red Selector Switch located on the front of the LCR and

remove the plate.

Rotate the Selector Switch counter-clockwise to the 6 o'clock (CALIBRATION/PROVER) position to permit access

to the LCR calibration menu.

STEP 2: CALIBRATION/PROVER Position

When the switch is rotated to the CALIBRATION/PROVER Position, the initial display is Prover
Quantity

.

Calibration - Screen 1, Prover Quantity

The units of measure are specified on the face label of the LCR. If a delivery is NOT active, press INCREASE to start

a new delivery. If a delivery is active, press INCREASE to stop the delivery and advance to Calibration - Screen 5,
Editable Prover Quantity

(described on Page 31).

SINGLE POINT RECALIBRATION/PROVING

000.000