beautypg.com

Lap pad calibration & set-up – Liquid Controls LCR Setup & Operation User Manual

Page 15

background image

15

Lap Pad Calibration & Set-up

STEP 4: SYSTEM CALIBRATION CONTINUED

DECIMAL^

This field allows you to choose between whole units, tenths and hundred ths of a unit. Press ENTER

and the cursor will drop to the bottom line. Use LLLLL to scroll between WHOLE, TENTHS, and HUNDR.

When the desired choice is displayed press ENTER and the cursor will move to the top line.

Press LLLLL and the cursor will move to:

RESIDUAL^

Volumes less than the least significant digit can be rounded into that digit or truncated (thrown away)

depending on the selection mode. Press ENTER and the cursor will drop to the bottom line. Press LLLLL to

scroll between ROUND and TRUNCATE. When the desired choice is displayed press ENTER and the

cursor will move to the top line.

Press LLLLL and the cursor will move to:

TEMP

This field displays the current temperature reading from the RTD temperature probe and allows one to

enter a corrected value from a Weights & Measures thermometer. If the LCR is not equipped with the

optional Electronic Temperature Volume Compensation (ETVC) Kit, this field will read "--".

To make a correction:

Run enough product through the meter to allow the temperature to st abilize. Compare the
TEMP

reading with the current Weights & Measures thermometer reading. If the readings are

different, the Weights & Measures reading should be entered. Press ENTER and the cursor

will drop to the bottom line. Enter the new value. The new value will overwrite the previous one

and the OFFSET field will be recalculated by the LCR. Offset adjustments are limited to ±0.3°C

(±0.54°F). Adjustments greater than this limit require replacement of the R TD probe. Press
ENTER

and the cursor will move to the top line.

Press LLLLL and the cursor will move to:

OFFSET

The offset represents the difference between the of ficial Weights & Measures reading and the LCR

RTD reading. The offset is automatically calculated if an entry is made in the previous field.

If a Weights & Measures thermometer reading was not entered in the previous field, subtract the value

listed in the TEMP field from the Weights & Measures measurement. Key in the difference in the offset

field. If the number is negative, use " *" on the Lap Pad to change the sign. Adjustment is limited to

±0.3°C (±0.54°F). Any adjustment larger than ±0.3°C (±0.54°F) generates a "RANGE ERROR" message

and requires replacement of the RTD probe.

Press LLLLL and the cursor will move to:

TEMP

25.53

System Calibration - Screen 3

RTD OFS

2.817

RTD SLP

0.024724

T TEMP^

DEG. F

OFFSET

0.00