Display values, Process data controller block – Hach-Lange SC 200 Profibus network card User Manual User Manual
Page 10

When the manual configuration is selected, the telegram data structure
can be configured by the user (refer to
Table 2 Profibus data telegram structure
Byte number
Data
Data type
1–2
Classified error
Integer (2 bytes)
3–4
Classified status
Integer (2 bytes)
5–8
Measurement 1
Floating (4 bytes)
9–12
Measurement 2
Floating (4 bytes)
13–16
Measurement 3
Floating (4 bytes)
Display values
The Profibus data block structure (
Profibus message data block
structure
) can replace sc probes without changes in the PLC configuration.
The primary value is always the measured value.
The secondary value, if not available, is filled with zero.
The tertiary value, if not available, is filled with zero.
Figure 2 Profibus message data block structure
Process data controller block
The data block for the sc controller is similar to the data block for sensors.
The structure of the sc controller data block is independent of the number
of connected sensors:
• sc controller_ERROR
• sc controller_STATUS
• Primary value
• Secondary value
• Tertiary value
Block 3 sc controller ERROR
and
Block 3 sc controller STATUS
show the
data definitions for error and status 1 in the sc controller.
Table 3 Block 3 sc controller ERROR
Bit
Error
Note
0
Sensor 1 communication
error
A communication error has occurred
between the sc controller and sensor 1, the
sensor could be disconnected.
1
Sensor 2 communication
error
A communication error has occurred
between the sc controller and sensor 2, the
sensor could be disconnected.
2–15 Not used
Table 4 Block 3 sc controller STATUS
Bit
Status1
Note
0
Sensor 1 installed First sensor has been installed to the sc controller. This
bit is set even if the sensor is disconnected after
installation.
1
Sensor 2 installed Second sensor has been installed to the sc controller.
This bit is set even if the sensor is disconnected after
installation.
2
Relay A on
3
Relay B on
10 English