Advanced operations – Hach-Lange SC 1000 User Manual
Page 86

84
Advanced Operations
Relation between input current and calculated concentration
shows the output value depending on the input current, the set low value and
the set high value with an input range of 0
–
20 mA.
SET HIGH VALUE
Default value: 20
Sets the value of the output value when the input current is 20 mA.
SET LOW VALUE
Default value: 0
Sets the value of the output value when the input current is 0 mA (0–20 mA scale) or 4 mA
(4–20 mA scale).
ON ERROR MODE
Default value: OFF
An error is reported when input current is out of range (this can be 0–20 mA or 4–20 mA).
When set to “OFF” no error will be reported even if the input current is out of range.
0 mA
Substitute value is 0 mA in case of error.
4 mA
Substitute value is 4 mA in case of error.
20 mA
Substitute value is 20 mA in case of error.
OFF
No substitute value is used to replace the measured value in case of an error.
CONCENTRATION
Displays the calculated concentration depending on the input current and the scaling which is set
in the SET LOW VALUE and SET HIGH VALUE menu.
LOG INTERVAL
Default value: 10 minutes
Sets the interval for logging the displayed value into the data logger.
Options: OFF, 5 minutes, 10 minutes, 15 minutes, 20 minutes, 30 minutes
VERSION
Displays the software version number
LOCATION
Displays the current location
sc1000 SETUP
CURRENT INPUTS
mA INPUT INT/EXT