Troubleshooting – Hach-Lange SC 1000 User Manual
Page 129

127
Troubleshooting
6. Cross-check the small input current.
7. Repeat the steps 1–6 until the input precision reaches a desired value.
TEST/MAINT
CURRENT INPUTS
mA INPUT INT/EXT
FUNCTION TEST
INPUT 1-4
INPUT CURRENT Displays the input current according to the current set to the corresponding channel.
SET FACTOR
Default value: 1
Adjusts the displayed input current with an factor.
SET OFFSET
Default value: 0
Adjusts the displayed input current with an offset.
OUTPUT MODE
Default value: HOLD
Decides how other members react if they try to read an input current value while the input
current card is in the function test. As the value of the input current in the function test is not
based on any measurements other members reading this value probably have to be informed
about this special situation. There are three settings: Hold, Active and Transfer.
HOLD
Another reading member does not use the current read value but uses the last value before the
output current card was in the function test mode.
ACTIVE
Another reading member uses the current value even if the output current card is in the function
test mode.
TRANSFER
The reading member uses an own substitute value for its own output value.