4 electrical data, Electrical data – Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual
Page 93
WQL-Cond
Technical data
189
ba75822e01
03/2011
14.4 Electrical data
Material
Metal parts coming into
contact with the sample:
– Shaft
– Sleeve
– Coupling ring
– Protective hood
– Cap
Stainless steel 1.4571*
Battery compartment
Brass, gold-plated
Screws
Stainless steel V2A
Housing of the control panel Brass, gold-plated
Key button
Silicone
Seals
FPM (Viton)
Shackle
Stainless steel V4A *
Shaft of the measuring cell
Epoxy
Conductivity electrodes
Graphite
Thermistor enclosure
Graphite
* Stainless steel can be susceptible to corrosion at chloride concentrations
of
≥
500 mg/l and more.
Guidelines
and norms used
EMC
EC directive 2004/108/EC
EN 61326
FCC Class A
Meter safety
EC directive 2006/95/EC
EN 61010-1
Climatic class
VDI/VDE 3540
Pressure resistance
EC directive 97/23/EC
IP protection class
EN 60529
Power
supply
Battery
Lithium thionyl chloride battery 3.6 V,
size AA, 2600 mAh
Operational life
At least 1 month, depending on measuring
interval
Electrical safety
Protective class
III
USB interface
Type
USB 1.1, galvanically isolated
Baud rate
38.400 (921.600 when reading in data)
Socket type
Mini USB
Cable length
max. 3 m