2 communication of the logger and pc program, 3 communication between pc program and sql server, Communication of the logger and pc program – Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual
Page 90: Communication between pc program and sql server, Section 13.3 c, Ommunication, Between, Program, Server, See section 13.3 c

What to do if...
WQL-Cond
186
ba75822e01
03/2011
13.2 Communication of the logger and PC program
Import of
measured values
is aborted
Data logging of the
logger aborted
13.3 Communication between PC program and SQL server
The program
cannot access the
server database
Cause
Remedy
– Communication problem
– Repeat the import process
– Use a shorter USB cable
– USB-Hub or USB connection at
the PC faulty or defective
– Use other USB connection or PC
Cause
Remedy
– Power supply interrupted
– Connect the logger to the PC
program and set up logging job.
– Make sure that the logger is
supplied with power (battery)
Cause
Remedy
– The SQL server is already
installed on the PC with a
different password
– In the Settings / Database menu,
click the User-defined option and
enter the password that was used
for the initial installation of the SQL
server.
– Set Windows authentication.
Note
The connection can be checked with the Test button. A message appears
informing you whether the connection was established.