Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual
Page 46

Logging
WQL-Cond
142
ba75822e01
03/2011
Temperature
compensation
The calculation of the temperature compensation is based on the preset 
reference temperature, 20 °C or 25 °C. The reference temperature is shown 
on the display.
Temperature
coefficient
The temperature coefficient for the linear temperature compensation is set to 
2.000 %/K in the factory. You can adjust this factor to meet your requirements 
in the range 0.000 ... 3.000 %/K. The factor is set in the menu for the setting 
of the temperature compensation.
TDS factor
The factor to calculate the total dissolved solids is set to 1.00 in the factory. 
You can adjust this factor to meet your requirements in the range of 0.40 to 
1.00. The factor is set in the menu for the parameter, TDS.
Time
Automatic time 
adjustment
z
 Yes (date/time for the logger is 
automatically synchronized with that 
of the PC)
z
No
(Enter the date/time for the logger)
Parameter
Mode
z
Conductivity 
[
μS/cm] / [mS/cm]
z
Salinity Sal [ ]
z
TDS 
[mg/l] / [g/l] with adjustable TDS 
factor in the range 0.40 ... 1.00
z
Resistivity 
[
Ω·cm] / [kΩ·cm] / [MΩ·cm]
Temperature 
compensation 
z
Nonlinear temperature 
compensation (nLF) according to 
EN 27 888
z
Linear temperature compensation 
(lin) with adjustable coefficients in 
the range 0.000 ... 2.000 ... 3.000 %/
K
z
No temperature compensation (off)
Reference 
temperature 
z
25 °C
z
20°C
Tab
Setting
Setting options
(default setting in bold)