Logging
WQL-Cond
142
ba75822e01
03/2011
Temperature
compensation
The calculation of the temperature compensation is based on the preset
reference temperature, 20 °C or 25 °C. The reference temperature is shown
on the display.
Temperature
coefficient
The temperature coefficient for the linear temperature compensation is set to
2.000 %/K in the factory. You can adjust this factor to meet your requirements
in the range 0.000 ... 3.000 %/K. The factor is set in the menu for the setting
of the temperature compensation.
TDS factor
The factor to calculate the total dissolved solids is set to 1.00 in the factory.
You can adjust this factor to meet your requirements in the range of 0.40 to
1.00. The factor is set in the menu for the parameter, TDS.
Time
Automatic time
adjustment
z
Yes (date/time for the logger is
automatically synchronized with that
of the PC)
z
No
(Enter the date/time for the logger)
Parameter
Mode
z
Conductivity
[
μS/cm] / [mS/cm]
z
Salinity Sal [ ]
z
TDS
[mg/l] / [g/l] with adjustable TDS
factor in the range 0.40 ... 1.00
z
Resistivity
[
Ω·cm] / [kΩ·cm] / [MΩ·cm]
Temperature
compensation
z
Nonlinear temperature
compensation (nLF) according to
EN 27 888
z
Linear temperature compensation
(lin) with adjustable coefficients in
the range 0.000 ... 2.000 ... 3.000 %/
K
z
No temperature compensation (off)
Reference
temperature
z
25 °C
z
20°C
Tab
Setting
Setting options
(default setting in bold)