14 Technical data
14.1 Measurement characteristics
General features
of the measuring
cell
Measuring principle
Four-electrodes measurement
Range of application
1 µS/cm ... 1 S/cm
Cell constant
0.475 cm
-1
±1.5 %
Temperature sensor
Integrated NTC 30 (30 k
: at 25 °C / 77 °F)
Temperature responding
behavior
t
99
(99 % of the final value display after)
< 20 s
Accuracy of the temperature
sensor
± 0.2 K
Measuring ranges,
resolution
(AutoRange
function:
the logger
automatically
selects the suitable
measuring range)
Measured parameter
Measuring range
Resolution
Conductivity [µS/cm]
0.0 ... 199.9
200 ... 1999
0.1
1
Conductivity [mS/cm]
2.00 ... 19.99
20.0 ... 199.9
200 ... 1000
0.01
0.1
1
Resistivity [Ohm*cm]
1.00 ... 19.99
20.0 ... 199.9
200 ... 1999
0.01
0.1
1
Resistivity [kOhm*cm]
2.00 ... 19.99
20.0 ... 199.9
200 ... 1999
0.01
0.1
1
Resistivity [MOhm*cm]
2.00 ... 19.99
0.01
Salinity
0.0 ... 70.0
according to the IOT
table
0.1
TDS
0 ... 199.9 mg/l
200 ... 1999 mg/l
2.00 ... 19.99 g/l
20.0 ... 199.9 g/l
1
0.01
0.1
Temperature [°C]
− 5.0 ... + 60.0
0.1
Temperature [°F]
+ 23.0 ... + 140.0
0.1
Accuracy of the
measuring tech-
nique in the range
of application
Measured parameter
Accuracy (± 1 digit)
Conductance
± 0.5 % of measured value
T [°C]
± 0.1