2 calibrating (automatic), Calibrating (automatic) – Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual
Page 33

WQL-Cond
Calibration in control standard (determining the cell constant)
129
ba75822e01
03/2011
6.2
Calibrating (automatic)
Preparatory
activities
Calibration
Note
We recommend to always calibrate with the protective hood screwed on so
the measuring cell is protected. Use a stand as necessary.
1
Connect the WQL-Cond logger to a USB port of your PC with the
USB cable.
2
Have the 0.01 mol/l KCl control standard ready.
3
In the Device menu, click the menu item, Calibrate.
The calibration window appears.
4
Immerse the logger in the 0.01 mol/l KCl control standard.
5
Use the Next button to start the measurement.