background image

Technical data

WQL-Cond

188

ba75822e01

03/2011

14.2 Application characteristics

14.3 General data

Logger function

Measuring interval

adjustable:
user-defined (1 s ... 24 h)
1 / 5 / 10 / 30 s
1 / 5 / 10 / 15 / 30 min
1 / 2 / 3 / 6 / 9 / 12 / 24 h

Capacity of the data
memory

600.000 measurement datasets

Measuring

medium

Allowed temperature range

0 ... 60 °C (32 ... 140 °F)

Allowed pH range

2 ... 12

Minimum depth of
immersion

60 mm

Pressure

resistance

IP 68 (1 x 10

6

Pa or 10 bar)

The WQL-Cond logger meets all requirements according to article 3(3) of
97/23/EC ("pressure equipment directive").

Storage

conditions

Recommended storing type

Dry, in air

Storage temperature

-25 ... 65 °C (-13 ... 149 °F)

Test certificates

CE

Dimensions

(in mm)

Weight

950 g

25.0 ± 0.1

6.5

422

Thread M6