Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual
Page 29

WQL-Cond
Setting the measuring mode
125
ba75822e01
03/2011
Application tips
Select the temperature compensations given in the table according to the
respective test sample:
Test sample
Temperature compensation
Display
Natural water
(ground water, surface
water, drinking water,
waste water)
nLF
according to EN 27 888
nLF
Other aqueous solutions
Set linear
temperature coefficient
0.000 ... 3.000 %/K
Linear
Salinity (seawater)
Automatic according to IOT
(International Oceanographic
Tables)
1
In the Device menu, select the menu item, Set.
2
The window for the selection of the measuring mode, temperature
compensation and reference temperature appears.
3
Select the required settings with a mouse click.
Clic Send settings.
A confirmation message appears: