Long term storage 626250, Sensor replacement, Charging dc batteries – YSI 5500D-1 User Manual
Page 99: Aquamanager database maintenance, Replacement parts & accessories, Appendices
YSI 5X00
196
Long Term Storage 626250
Long term storage is defined as a period of time > 30 days.
The 626250 sensor MUST BE STORED IN A MOIST ENVIRONMENT, i.e., either
in water or in water-saturated air. If the sensor cap is allowed to dry out by exposure
to ambient air, it is likely to drift slightly when next used.
1. Moisten the sponge in the red protective plastic cap that was provided with
the probe and place it over the sensor with the sensor cap installed.
2. Inspect the sponge every 30 days to make sure it is still moist. If you no lon-
ger have the red protective cap, then moisten the sponge in the calibration/
transport sleeve and place this over the probe.
3. Alternatively, you can place the probe with sensor cap directly in a beaker
or other container of water, making sure that the water does not evaporate
over time.
Sensor Replacement
The 5422 DO sensor is user replaceable. Refer to the instruction sheet provided
with the replacement sensor for detailed instructions.
Charging DC batteries
The 5X00 does not charge batteries. Quality assurance maintenance procedures
should be established if batteries are to be fully powered when used as back-up
power source
.
AquaManager Database Maintenance
See AquaManager Database Maintenance - page 182 and AquaManager online Help
for additional information.
Replacement Parts & Accessories
See www.YSI.com for replacement part and accessory information and Appendix
5 of this manual - page 232.
Maintenance
Appendices
-Appendix 1 - Unpacking and Inspection
-Appendix 2 - Troubleshooting
-Appendix 3 - Menu Maps
-Appendix 4 - Declaration of Conformity-
-Appendix 5 - Parts and Accessories
-Appendix 6 - SMS Messaging
-Appendix 7 - Alpha “A” Curve
-Appendix 8 - Warranty
-Appendix 9 - Contact Us
-Appendix 10 - Recycling