Component and system, Non input), Specifications – YSI 5500D-1 User Manual
Page 9
YSI 5X00
17
YSI 5X00
16
Component and System
(non input)
Specifications
Hardware
System
System Feature
Specification
Display/LEDs
backlit graphic LCD (320X240)
with timer
readings updated 2 times a
second
diagnostic troubleshooting
LEDs
Outputs
8 on-board relays
240VAC@ 5AMPs switching
capacity; 110VAC@10AMPS;
type: SPDT
Timer Systems
4 timers
10 daily events per timer system
4 feed timers
includes sensor & FCR control
Power: There are two 5X00 versions available. The DC version can only be powered by
DC power. The AC version can be powered by AC or DC power.
Power for DC
Version:
external 12VDC power source
9-16.5VDC; 800 mA
Power for AC
Version:
internal AC switching power
supply
100-240VAC; 50-60Hz; 300 mA
Both AC and DC
versions:
lithium battery (for real time
clock)
2032 lithium ion
Electronics
ARM7 embedded microproces-
sor
DSP (digital signal processing)
Digital processing of 5X00
information ensures accurate and
repeatable readings.
Introduction
Firmware
Feature
Specification
Flash Memory
Firmware field upgradable with
local PC connection
Data Logging
non-volatile data log memory
storage of 4750 records
sensor log (2000 records)
event log (1250 records)
calibration log (500 records)
Configuration
log (1000 records)
Security
password protection
Data filtering
reads 2-3 times per second; read-
ings are averaged
Communication
Feature
Specification
RS232
up to 100 ft (30m)
OR
Optional serial to ethernet
device
remote TCP/IP PC access and/or
email alarming
Instrument to Instrument Networking
Feature
Specification
Network 2 to 32 isolated 5X00
and/or 5200A instruments
RS485 up to 4000 ft (1220m)
entire network distance
AquaManager
Feature
Other PC Requirements
Runs on Windows 2000/XP/
Vista/Windows 7 operating
systems 32-bit and 64-bit
PC capable of running Windows
2000
VGA display card
CD-ROM Drive
Total RAM Installed 16
Megs or greater
100 MB free hard-disk space
Introduction