Sensors, Calibrate – YSI 5500D-1 User Manual
Page 84
YSI 5X00
167
YSI 5X00
166
Sensors
Menus →Timers→Feed Smart Timer→Feeder→Sensors
→Disabled/Enabled
→DO Low/High/DO % Change
→Temp Low/High/Temp % Change
Changes in water conditions affect the amount of food fish eat which in turn affects
fish growth. With the sensor feature enabled, the feed timer system will verify that
DO and temperature values are within user configured ranges. This verification is
done immediately prior to starting each feed duration in a feed cycle. If values are
reported that fall outside acceptable configured ranges, the feed timer relay energizes
for a shorter period (user configured percentage) thereby reducing the amount of
food that is dispensed. Feed timer sensor ranges are independent of sensor setup
configuration ranges. For timed feed cycles, feed interval for the entire feed cycle
is determined by the duration of the first feed.
Example - The feed timer is configured as follows: sensor temperature low value
is 25.1oC, feed timer temperature change is 9%, timed feed mode, six feeds, daily
weight 2 units start time 8A, end time 9A, six 9.999 second feeds of .3333 units
each to occur at 9 minute 50 second intervals without sensor reduction. Prior
to 8A feed cycle, the start temperature value is reading 24.9oC. Feed duration of
9.999 seconds will be reduced by approximately 1 second (9%) because of the “out
of range” temperature value.
Notes:
- 5400 - Feed Timer temperature sensor value is always taken from the DO
system configured temperature source.
- Feed Timer temperature source is always taken from the ODO temperature
sensor regardless if temperature system is configured for an aux temperature
sensor.
DO Low/Temp Low
Enter the lower limit of the sensor range where no feed reduction will take place.
5400
5500
Configuring the 5X00
DO High/Temp High
Enter the high limit of the sensor range where no feed reduction will take place.
DO Change/Temp Change
Enter the percent reduction in feed weight that will occur for “out of range” enabled
feed timer sensors. Results from all out of range sensors are added. To disable a
sensor value from affecting feed duration configure 0.0% as the % change for that
sensor.
Calibrate
Menus →Timers→Feed Smart
tm
Timer→Feeder→Calibrate
Calibrating the feeder is essential for proper operation of the feed timer system.
Feed timer calibration is required to establish the amount of food dispensed over
a certain period of time. The 5X00 will base feed cycle duration times on the last
saved calibration value.
Calibrate Feed Smart
tm
Timer
1. Install and wire feed timer device. See wire relays - page 56.
2. Enable Feed Timer.
Menus → Timers→Feed Smart
tm
Timer →Feed Timer →Enabled
3. Assign feed timer relay.
Menus → Timers→Feed Smart
tm
Timer →Relay
4. Calibrate feeder.
Menus → Timers→Feed Smart
tm
Timer →Calibrate
5. At numeric keypad, enter test run time the feeder relay will energize.
Configuring the 5X00