Probe name, Input type – YSI 5500D-1 User Manual
Page 53
YSI 5X00
105
YSI 5X00
104
5400 Configure Aux 5 menu
5500 Configure Aux 5 menu
(no Temp Source menu)
(
Temp Source menu will be available when Tempera-
ture
Input
Type
configured
Probe Name
Menus → System → Configure Aux →Aux 1-6→Probe Name
Use the alpha/numeric keypad to name aux inputs. Configured name(s) appear
throughout menus and at Run Screen - figure 4.7- page 108.
Note:
- Aux names can also be configured at Menus → System → Name Devices/
Sensors.
Input Type
Menus → System → Configure Aux →Aux 1-6→Input Type
→Digital
→0-1v
→0-5v
→4/20ma
→Aux 3-6→Input Type
→Temperature
Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup
menu selections are based on the aux input type configured at system menu. Input
types are discussed in the following sections.
Configuring the 5X00
Digital
Menus → System → Configure Aux →Aux 1-6→Input Type→Digital
Use digital input when wiring an external device to control or alarm when the
contacts of an external switch open or close. A common application would be to
detect high or low water levels using a float switch.
Notes:
- Suffix, Min, Max, and Display Format are not available for digital inputs.
- In the 5X00 sensor log, aux digital “closed” is logged as Aux 0.00. Aux digital
“open” is logged as Aux 1.00.
1. Wire aux digital devices - page 59.
2. Enter Aux menu
Menus → System → Configure Aux →Aux 1-6
3.
Use
▲
and
▼
to scroll and highlight submenus.
4.
Press
to select.
5. Configure digital input type and name.
6. Press
to save configuration and return to Run Screen.
7. Go to Sensor Setup Menus → Sensor Setup Aux 1-6 - page 147 to
enable and configure control/alarm on condition, control/alarm relay,
and hold off time.
8. Press
to save configuration.
Analog Non Temperature
0-1VDC and 0-5VDC and 4-20 mA
Menus → System → Configure Aux →Aux 3-6→Input Type
→0-1V
→0-5V
→4/20ma
Select the analog input type based on the signal being sent from the device.
Configuring the 5X00