YSI 5500D-1 User Manual
Page 55
YSI 5X00
109
YSI 5X00
108
figure 4.7
Notes:
- Values displayed at the run screen are based on display format resolution
configuration. Example - if the display format is 0.0 and the min value input
is 1.88, then 1.9 will be displayed at the run screen.
- 5X00 aux analog (non temp) sensor log records 0.00 resolution regardless of
the selected resolution in the aux analog display format configuration.
Directions to configure Aux analog (non temp) Configure Aux menu:
1. Wire aux analog devices - page 59.
2. Enter Aux menu
Menus → System → Configure Aux →Aux 1-6
3.
Use
▲
and
▼
to scroll and highlight submenus.
4.
Press
to select.
5. Press
to save configuration.
6. Go to Sensor Setup Menus → Sensor Setup Aux 1-6 - page 149. to en-
able and configure set points, control, alarm, relays, and hold off time.
7. Press
to save configuration.
Note:
- The alarm low value must be ≥ the min value. The alarm high value must be
≤
the max value.
Analog Temperature (Aux 3, 4, 5 and 6)
Menus → System → Configure Aux →Aux 3-6→Input Type→Temperature
Aux 3, 4, 5 and 6 can be wired and configured for a temperature sensor. Temperature
systems are configured differently for 5400 and 5500D instruments. The differences
are explained below.
5400 ~ Aux configured temperature systems can be configured as stand alone
5400
Configuring the 5X00
independent temperature systems. In addition, Aux temperature systems can be
configured as the temperature source for any of the four DO systems. Configure the
DO temperature source in the DO sensor set up menu - see page 154.
5500D-01, 5500D-02 and 5500D-04 ~ Aux configured temperature systems can
only be configured as stand alone independent temperature systems and cannot be
assigned as a temperature compensation source for any DO system. The DO system
temperature compensation value is always taken from the temperature sensor in the
ODO probe. Temperature systems can be configured using the ODO temperature
value or the Aux 3, 4, 5 and/or 6 temperature value. Configure the temperature
system source at the Menus → System → Configure Aux →Temp Source menu.
Configure “Aux Port” if the temperature system is to be driven by the input value
of a temperature sensor wired to aux 3, 4, 5 or 6. Configure “ODO Probe” if the
temperature system is to be driven by the ODO probe temperature sensor value.
Directions to configure Aux temperature Configure Aux menu:
1. Wire temperature sensors - page 54.
2. Enter Aux menu
Menus → System → Configure Aux →Aux 3-6
3. Configure temperature input type and name.
4.
Use
▲
▼
to scroll and highlight submenus.
5.
Press
to select.
6. Press
to save settings and return to Run Screen.
7. Go to Sensor Setup Menus → Sensor Setup Temperature for Aux 3-6
and ODO - page 150 to configure temperature system.
8. Press
to save settings and return to Run Screen.
Note:
- The Sensor Setup menu - page 137 is automatically enabled when an aux
channel is configured for temperature.
5500
5500
Configuring the 5X00