Run screen – YSI 5500D-1 User Manual
Page 39
YSI 5X00
77
YSI 5X00
76
Softkey label
Function
Exit menu return to Run Screen saving
menu configuration changes.
Exit menu (to previous menu) without sav-
ing configuration changes.
Back to previous menu. Changes to configu-
rations are saved at return to Run Screen.
Prompt to take next action. i.e. enter user
value at calibration menus.
Save alpha and numeric configurations and
calibrations.
Disable/Enable menu option. i.e. buzzer.
Scroll left and right in alpha and numeric
keypad screens.
To acknowledge and save sensor calibration.
Prompt to confirm “Are you sure?” message
when performing certain functions. i.e. reset
to factory and clear data log menus.
Digital input control configuration from
Sensor Setup Menu.
table 4.1
Run Screen
The Run Screen displays measurements and inputs from enabled sensors and
conditions - figure 4.2. At power up, the Run Screen is displayed immediately after
the YSI splash screen (the YSI splash screen appears for approximately 2 seconds).
Icons displayed at the Run Screen represent individual and systemwide conditions.
See Run Screen Icons - page 78.
Run Screen and sensor display formats are user selectable. See Display Settings - page
115 and Sensor Setup Menus - page 137.
Notes:
- The Run Screen must be displayed during normal operation.
- All alarm functions (5X00 email alarms, energizing of enabled alarm relays,
alarm icons, and buzzer) are suspended when in 5X00 instrument menus.
Alarm systems are reset when menus are exited to Run Screen.
Configuring the 5X00
- Always manually poll the 5X00 after uploading configuration changes and
verify the uploaded information is retained. When configuration changes
are uploaded using AquaManager at the same time configuration changes are
being made at the instrument, the AquaManager uploads may not be saved
depending on the timing of keystrokes at the 5X00.
- The Run Screen must be displayed during normal operation. If the 5X00
is not at the run screen during AquaManager uploading, pressing
at
the 5X00 will only save certain configuration changes.
- At power up, 5X00 systems have a 30 second hold off allowing sensor values
to stabilize before enabling any control and/or alarm devices.
Figure 4.2
Configuring the 5X00