beautypg.com

7 data image in vegacom 557 – VEGA VEGACOM 557 Profibus DP User Manual

Page 28

background image

28

VEGACOM 557 Profibus DP

Data image in VEGACOM 557

7 Data image in VEGACOM 557

all measured values
of VEGAMET #2

Grouping of measured values acc. to VEGAMET

addresses (here VEGAMET 513)

all measured values
of VEGAMET #15

all measured values
of VEGAMET #1

all measured values

of VEGAMET #3

VEGACOM 557 collects the measured values

of the VEGA signal conditioning instruments

VEGAMET series 500 and 600 (via DISBUS)

or VEGALOG 571 (via LOGBUS) and puts
them in the temporary memory for collection

via the Pr ofibus DP. The method of storing

the measur ed values (for the higher -priority

processing system) in the VEGACOM 557

temporary memory differs depending on the
selected configuration. It depends on

whether VEGACOM 557 is connected to the

DISBUS or the LOGBUS, and if DISBUS, also

on the connected instrument type. The alloca-

tion can, however, also be influenced via
appropriate inquiry protocols.

Note

On the VEGAMET 513, 514, 515 and series

600 signal conditioning instruments, as well
as on the VEGALOG 571 processing system,

the individual configuration of the PC/DCS

outputs with the adjustment software VVO is

possible (see chapter “4 Measurement loop
in VEGAMET/VEGALOG “).

On VEGAMET 513, 514, 515 and series 600

signal conditioning instruments as well as the

VEGALOG 571 processing system, it is
additionally possible to enquire the conditions

of the contact inputs/ and outputs via VEGA-

COM 557.

7.1 Image of measured value when

connecting to DISBUS

UP to 15 bus participants (VEGAMET) can

be connected to the DISBUS, whereby each

participant delivers max. 7 PC/DCS values.

The bus addresses 1 … 15 are configured
on the VEGAMET instruments, bus address 0

is not permitted. As already mentioned, the

storage of the measured values for the

VEGAMET signal conditioning instruments

can be influenced by the user . For this, the
operating mode of VEGACOM 557 and the

enquiry telegram (control command) of the

Profibus-Master are decisive.

Image of measured value DISBUS in
mode 50E2A

In this operating mode of VEGACOM 557, the

PC/DCS values are allocated as blocks relat-
ing to the VEGAMET bus address and the

PC/DCS output number .

A memory area of 16 PC/DCS values is re-

served for each VEGAMET . The allocation of

the individual PC/DCS values in blocks also
depends on the enquiry telegram (control

command) of the Pr ofibus-Master. Ther e are

two control commands to which the VEGA-

COM 557 reacts. The two control commands

0xCC or 0xCB are described in detail in the
following.

Control command 0xCC (DISBUS)

In case of an enquiry with this control com-
mand, the PC/DCS values will be grouped in

ascending sequence acc. to the VEGAMET

addresses.

In total, there are 32 blocks with 8 PC/DCS

values each.