M&C TechGroup CSS-V2 Operator's manual User Manual

Page 27

background image

4-1.2-ME

Gas sampling and gas conditioning technology

27




Sample line to the
analyser clogged or
squeezed;




Pollution of the
diaphragm pump;


No gas flow?

Loosen the outlet hose on the analyser side and
check on the threaded hose coupling whether
sample gas flows;
Sample gas does not flow?
Clean the clogged line or replace it;
Sample gas flows?
ok?

Loosen the piping on the pump head and examine
it (see 15.2);
ok?
If necessary, clean the pump;
Ok?

Temp.
<2°C

Cooler switches off
the sample gas pump
automatically;
Cooler defective;

Ambient temperature too low;
ok?

Return the instrument for repair to M&C.