Receipt of marchendise and storage, Installation instructions – M&C TechGroup CSS-V2 Operator's manual User Manual
Page 14

14
Gas sampling and gas conditioning technology
4-1.2-ME
10
RECEIPT OF MARCHENDISE AND STORAGE
The gas sampling system CSS-V.. is a completely pre-installed unit. The scope of delivery includes
furthermore:
25 Filter elements (1 pckg.)
1 Connection cable
1 6-pole connection box
1 Operating Manual
Please take the CSS-V.. and possible special accessories carefully out of the packaging material
immediately after arrival, and compare the goods with the items listed on the delivery note;
Check the goods for any damage caused during delivery and, if necessary, notify your transport
insurance company without delay of any damage discovered.
N O T E !
The gas conditioning system should be stored in a protected frost-free area
!
11
INSTALLATION INSTRUCTIONS
N O T E !
In order to ensure that the case is safe and stable when being used, it should
be placed on a horizontal surface free from vibrations. Only then, the perfect
functioning of the separation and drainage procedures of the condensate
inside the heat exchanger of the cooler will only be guaranteed.
The gas conditioning system should be kept away from heat sources and be
freely ventilated in order to prevent an accumulation of heat.
When the equipment is being used outside, ample protection against the
effects of direct sunlight and dampness must be provided. In winter, the
equipment must only be used in frost-free areas; pay attention to the type of
the equipment protection.
Please avoid temperature variations, strong airflow as well as aggressive
atmospheres at the place of installation.
In order to guarantee the operational safety of the gas conditioning system
and the downstream analysers, and to avoid false alarms, the gas
conditioning unit should not be used at temperatures other than those
specified. Furthermore, it must be protected against dust deposits and
penetrating dust.
It is of great importance that the analysers connected downstream are used
at temperatures well above the specified gas outlet dew point of +5°C. This
prevents the gas in the connector lines to the analysers from subsequent
condensing.
Unheated gas sample lines must be installed with slope up to the cooler.