General troubles during operation – Olympus IPLEX MX II User Manual
Page 84
78
Chapter 8 Troubleshooting
General troubles during operation
Problem
Possible cause
Remedy
The illumination
does not light.
The LIGHT button (
) is not on.
Turn on the LIGHT button (
).
The LEDs on the distal end of the
insertion tube are inoperable.
Return the unit to Olympus for repair
investigation.
The illumination
is dim.
The distal end of the insertion tube is
dirty.
Wipe with a clean piece of gauze or a
cotton swab.
Extended use under high ambient
temperatures causes decreased light
emissions.
Return the unit to Olympus for repair
investigation.
The side view
mirror adapter
cannot be
attached to the
insertion tube.
(for IV86)
Foreign object in the distal end of the
insertion tube (closer to the tip than
the side view mirror adapter lock
groove) or inside the side view mirror
adapter.
Wipe them with a clean piece of
gauze or cotton swab.
The side view mirror adapter in use is
not a designated optical adapter for
this system.
Use a designated side view mirror
adapter.
You are not using the proper
installation procedure.
Redo it correctly (See “Attaching and
detaching the side view mirror
adapter” (page 28).)
The side view
mirror adapter
cannot be
removed from
the insertion
tube.
The nut of the optical adapter is
rotated in the wrong direction.
Redo it correctly (See “Attaching and
detaching the side view mirror
adapter” (page 28).)
The system
cannot be
turned on.
The AC adapter or the battery is not
connected.
Connect the AC adapter or the battery
correctly. (See “3.3 Preparing the
power supply” (page 22).)
The [POWER] button is not set to on.
Set the power button (
) of the main
unit to on.
A non-designated AC adapter or
battery is connected.
Use the designated AC adapter or
battery.
The system
cannot be
turned off.
The main unit is damaged.
Remove the AC adapter or the battery
from the main unit to turn off the
power.