Olympus IPLEX MX II User Manual
Page 83
77
Chapter 8 Troubleshooting
Self-check
triggers during
use and
displays a
message (and
an alarm
sounds).
HIGH
TEMPERATURE
(MAIN UNIT).
PLEASE TURN
OFF THE
POWER.
The message prompts you
to stop operations because
the inside of the main unit
was too hot and triggered
the self-check function.
Terminate inspection, allow
the instrument to cool down,
and then turn power back on
again.
The angulation
operation is
extremely
heavy.
OVER CURRENT
(ARTICULATION
MOTOR).
PLEASE TURN
OFF THE
POWER.
The insertion tube is
overloaded and the self-
check function was
triggered to prompt
termination of inspection.
Extend the insertion tube as
much as possible to reduce
the loop amount, and turn
power back on without
operating the [ANGLE]
joystick.
Problem
Display
Possible cause
Remedy