Chapter 4 Basic operations
Date and time settings
Set the current date and time before using the system for the first time. For more
information, see “Setting the date and time” (page 60).
Checking the illumination lighting
WARNING
• The Illumination emitted from the insertion tube distal end may heat and ignite a nearby
object. Be sure to set the illumination light turn off the LIGHT button (
) on the front panel of
the main unit when not using the instrument.
• Do not allow the light emitted from the insertion tube’s distal end enter your eye directly.
Otherwise, an eye injury may result.
1
Confirm that the light indicator is shown on the screen. If it is not,
press the LIGHT button (
) on the front panel to make it appear.
2
Verify the light is emitted from the insertion tube to ensure that the
light source is turned on.
White balance adjustment
Adjust the white balance of the insertion tube as required by following the
procedure described in “Live screen/frozen screen menu display and functions”
in “WHITE BALANCE” (page 57).
Checking the angulation functions
1
Straighten the insertion tube.
2
Manipulate the [ANGLE] joystick slowly and ensure that the
angulation section moves smoothly.
CAUTION
• Repair is required if the joystick shaft is bent or if the knob is detached. Contact Olympus.
NOTE
• The angulation movement is interlocked with the angle and direction of the [ANGLE] joystick
movements.