beautypg.com

RIGOL DS1000D User Manual

Page 82

background image

RIGOL Command Systems

Programming Guide for DS1000E, DS1000D Series

2-70

1. :WAVeform:DATA?

Command Format:
:WAVeform:DATA? []

Function Explanation:
The command reads the waveform data in the specified source. could
be CHANnel1, CHANnel2, DIGital, MATH or FFT.

Returned Format:
The query returns 1024 data.



2. :WAVeform:POINts:MODE

Command Format:
:WAVeform:POINts:MODE
:WAVeform:POINts:MODE?

Function Explanation:
This command sets the mode of waveform points. can be:
NORMal, MAXimum or RAW.


Data points returned by :WAVeform:DATA? in different modes:

NORmal

RAW

MAX

Normal Memory Long Memory

In RUN state,
MAX is the
same with
NORMal; in
STOP
state,
MAX is the
same with
RAW.

MATH

600

600

600

FFT

512

512

512

CHx

600

8192(8k)

524288(512k)

Half-
Channel
CHx*

600

16384(16k)

1048576(1M)

DIGITAL 600

16384(16k)

1048576(1M)

*Note: Half-channel means only one channel is open and the MATH is closed.

Returned Format:
The query returns NORMal, MAXimum or RAW.