RIGOL Command Systems

Programming Guide for DS1000E, DS1000D Series

2-56

Function Explanation:
The commands set and query the vertical offset. could be 1 or 2.
When Scale≥250mV, the range of is -40V~+40V;
When Scale<250mV, the range of is -2V~+2V.

Returned Format:
The query returns the setting value of .

Example:
:CHAN2:OFFS 20 Setup the vertical offset of channel 2 as 20V.
:CHAN2:OFFS?

The query returns 2.000e+01.


6. :CHANnel:PROBe

Command Format:
:CHANnel:PROBe
:CHANnel:PROBe?

Function Explanation:
The commands set and query the attenuation factor of the probe. could be
1 or 2; could be 1, 5, 10, 50, 100, 500 or 1000.

Returned Format:
The query returns the setting value of .

Example:
:CHAN2:PROB 10 Setup the attenuation ratio of channel 2 as 10X.
:CHAN2:PROB?

The query returns 1.000e+01.



7. :CHANnel:SCALe

Command Format:
:CHANnel:SCALe
:CHANnel:SCALe?

Function Explanation:
The commands set and query the vertical scale of waveform magnified by the