Trigger:alternation:timescale, Trigger:alternation:timeoffset – RIGOL DS1000D User Manual
Page 52
Programming Guide for DS1000E, DS1000D Series
2-40
3. :TRIGger:ALTernation:TimeSCALe
Command Format:
:TRIGger:ALTernation:TimeSCALe
:TRIGger:ALTernation:TimeSCALe?
Function Explanation:
The commands set and query the timebase of the current channel. The range of
Returned Format:
The query returns the setting value of the timebase in s.
Example:
:TRIG:ALT:TSCAL 0.001
Setup the timebase as 1ms.
:TRIG:ALT:TSCAL?
The query returns 1.000e-03.
4. :TRIGger:ALTernation:TimeOFFSet
Command Format:
:TRIGger:ALTernation:TimeOFFSet
:TRIGger:ALTernation:TimeOFFSet?
Function Explanation:
The commands set and query the horizontal timebase offset of the current
channel. Thereinto,
In NORMAL mode, the range of
In STOP mode, the range of
In ROLL mode, the range of
Scale indicates the current horizontal scalre, the unit is s/div.
Returned Format:
The query returns the setting value of offset in s.
Example:
:TRIG:ALT:TOFFS 0.0002 Setup the horizontal timebase offset as 200us.
:TRIG:ALT:TOFFS?
The query returns 2.000e-04.