3. :TRIGger:SLOPe:MODE

Command Format:
:TRIGger:SLOPe:MODE
:TRIGger:SLOPe:MODE?

Function Explanation:
The commands set and query the slope condition. could be
+GREaterthan (positive slope greater than), +LESSthan (positive slope less
than), + EQUal (positive slope equals to), -GREaterthan (negative slope greater
than), -LESSthan (negative slope less than) or –EQUal (negative slope equals
to).

Returned Format:
The query returns +GREATER THAN, +LESS THAN, +EQUAL, -GREATER THAN,
-LESS THAN 或-EQUAL.

Example:
:TRIG:SLOP:MODE +GRE Setup the slope condition as +GREaterthan.
:TRIG:SLOPe:MODE?

The query returns +GREATER THAN.


4. :TRIGger:SLOPe:WINDow

Command Format:
:TRIGger:SLOPe:WINDow
:TRIGger:SLOPe:WINDow?

Function Explanation:
The commands set and query the type of trigger level which can be adjusted by

.

When the slope condition is +GREaterthan, +LESSthan or + EQUal,
could be PA (rising edge Level A), PB (rising edge Level B) or PAB (rising edge
Level AB);
When the slope condition is -GREaterthan, -LESSthan or –EQUal, could
be NA (falling edge Level A), NB (falling edge LevelB) or NAB (falling edge
LevelAB).

Returned Format: