beautypg.com

Duration trigger, Duration trigger -37, Trigger:duration:pattern – RIGOL DS1000D User Manual

Page 49: Trigger:duration:time

background image

Command Systems

RIGOL

Programming Guide for DS1000E, DS1000D Series

2-37

DURation Trigger

1. :TRIGger:DURation:PATTern

Command Format:
:TRIGger:DURation:PATTern ,
:TRIGger:DURation:PATTern?

Function Explanation:
The commands set and query the type of pattern for digital signal. Thereinto,
When is a 16-bit integer without sign, high is 1, and low is o;
When is a 16-bit integer without sign, enable is 1, and X is 0.

Returned Format:
The query returns the setting value of , in turns.

Example:
:TRIG:DUR:PATT 65535,655535 Setup the type of pattern.
:TRIG:DUR:PATT?

The query returns 65535,65535.


2. :TRIGger:DURation:TIME

Command Format:
:TRIGger:DURation:TIME

of


Example:
:TRIG:DUR:TIME 0.05 Setup the limiting time of duration trigger as 50ms.
:TRIG:DUR:TIME?

The query returns 5.00e-02.

This manual is related to the following products: