Channel:filter, Channel:memorydepth – RIGOL DS1000D User Manual

Page 69

background image

Command Systems

RIGOL

Programming Guide for DS1000E, DS1000D Series

2-57

oscilloscope. could be 1 or 2.
When the Probe is set to 1X, the range of is 2mV ~ 10V;
When the Probe is set to 5X, the range of is 10mV ~50V;
When the Probe is set to 10X, the range of is 20mV ~ 100V;

When the Probe is set to 50X, the range of is 100mV ~ 500V;
When the Probe is set to 100X, the range of is 200mV ~ 1000V;
When the Probe is set to 500X, the range of is 1V ~5000V;
When the Probe is set to 1000X, the range of is 2V~ 10000V.

Returned Format:
The query returns the setting value of in V.

Example:
:CHAN2:PROB 10 Setup the attenuation ratio of the probe as 10X.
:CHAN2:SCAL 20

Setup the vertical scale as 20V.

:CHAN2:SCAL?

The query returns 2.000e+01.


8. :CHANnel:FILTer

Command Format:
:CHANnel:FILTer {ON|OFF}
:CHANnel:FILTer?

Function Explanation:
The commands set and query the On/Off state of the filter. could be 1 or 2.

Returned Format:
The query returns ON or OFF.

Example:
:CHAN2:FILT OFF Turn off the filter of Channel 2.
:CHAN2:FILT?

The query returns OFF.



9. :CHANnel:MEMoryDepth?

Command Format:
:CHANnel:MEMoryDepth?